If not in use, adjust the range of the pipette gun to the maximum scale, so that the spring is in a relaxed state to protect the spring.
It is best to clean the pipette regularly, you can use soapy water or 60% isopropanol, then use distilled water to clean it, and let it dry naturally.
Before autoclaving, make sure that the pipette can adapt to high temperatures.
Calibration can be performed by repeating the method of weighing distilled water several times in an environment of 20-25 degrees.
Check if there is any leakage when using it. In the method, suck the liquid and place it vertically in the air for a few seconds to see if the liquid level drops. If the liquid leaks, the reasons are roughly as follows:
1. Whether the tip of the gun matches;
2. Whether the spring piston is normal;
3. If it is a volatile liquid (as many organic solvents are), it may be a problem of saturated vapor pressure. The liquid can be pipetted several times before pipetting.